86
by coolant and swarf on the tool.
Also check for mechanical movement of the tool/tip insert
or NC1 system.
This is a reset condition.
Message FALSE TRIGGER
Meaning The beam has been cut several times during a
measurement move. This may be caused by coolant
triggering the beam.
Action Identify and correct the cause of the false triggers. The
default number of retries allowed is set to 1. If necessary,
change this number in subroutine L9760 (for details, see
“Setting-data subroutine L9760” on page 13).
This is a reset condition.
Message OUT OF TOLERANCE
Meaning 1 Subroutine L9862: The measured length or diameter of
the tool is out of tolerance (either a positive or negative
limit is exceeded). This may be caused by a broken tool
or the tool being pulled out of its holder.
Action 1 Reset and replace the tool, or adjust the tool then reset it.
Meaning 2 Subroutine L9861: The temperature compensation drift
values have exceeded the allowable tolerance.
Action 2 Investigate the cause of excessive movement. The zero
offset registers used for temperature compensation may
have been wrongly adjusted since storing reference
values.
This is a reset condition.
Error messages and alarms

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Renishaw NC1 and is the answer not in the manual?

Renishaw NC1 Specifications

General IconGeneral
BrandRenishaw
ModelNC1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals