− Feed volume: max. 2 x 50 ml
3.10 Feed Grain Size
The feed grain size depends on the sample material and on the device configuration and
setting.
− Feed grain size: ≤ 5 mm
3.11 Cooling
Please refer to chapter "Commissioning of the Cooling" for detailed information on the
commissioning of the internal and external cooling.
3.11.1 Internal
There is a coolant tank at the backside of the device which must be filled with coolant before
putting into operation.
− Fill volume: ~ 600 ml
− Coolant: clean, lime-free water plus coolant additive
3.11.2 External
An additional cooling can be connected at the back of the device.
− Maximum pressure: < 6 bar
− Minimum temperature of the cooling liquid: > 5 °C
NOTICE Only clean, lime-free water with a coolant additive (article no. 02.362.0027) is
permitted as cooling liquid. The temperature must not fall below the minimum value, as
otherwise the sealing gaskets of the cooling circuit can be damaged.