Packaging, Transport and Installation
17
NOTICE
N7.0021
Ambient temperature
Temperatures outside the permitted range
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
Do not exceed or fall below the permitted temperature range (5 °C to
40 °C ambient temperature) of the device.
Installation height: max. 2000 m above sea level
Ambient temperature: 5 °C 40 °C
Maximum relative humidity < 80 % (at ambient temperatures 31 °C)
For ambient temperatures U
T
between 31 °C and 40 °C, the maximum relative humidity value L
F
linearly decreases according to L
F
= (U
T
55) / 0.3:
Ambient temperature
Max. rel. humidity
31 °C
80 %
33 °C
73.3 %
35 °C
66.7 %
37 °C
60 %
39 °C
53.3 %
40 °C
50 %
NOTICE
N8.0015
Humidity
High relative humidity
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
The relative humidity in the vicinity of the device should be kept as
low as possible.