Cleaning, Wear and Maintenance
66
Fig. 40: Removing the lid (left), draining the coolant and rinsing the system (right)
Fill 600 ml of clean, lime-free water into the coolant tank. The water is circulated through the
system and drained again via the hose. NOTICE Only use lukewarm water for rinsing. Do
not use boiling water and do not add any descaling agents!
Repeat this process approx. three to five times, until the turbidity of the water disappears and
the clear water flows from the cooling system into the collecting vessel.
Deactivate the cleaning mode as soon as the cooling system is emptied.
Remove the hose from the outlet socket (SU).
A residual cloudiness of the cooling liquid can persist, even after repeated rinsing. This,
however, is uncritical.
Fig. 41: Removing the pipe
Remove and clean the pipe (TR).