EasyManuals Logo

Rigol MSO5074 User Manual

Rigol MSO5074
372 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
RIGOL Chapter 1 Quick Start
1-8 MSO5000 User Guide
Function Inspection
1. Press Default on the front panel, then a prompt message "Restore default
settings?" is displayed. Press OK or tap OK to restore the instrument to its
factory default settings.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in Figure 1-8 below.
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe, as shown in Figure 1-8.
Figure 1-8 To Use the Compensation Signal
4. Set the probe attenuation to 10X, and then press AUTO.
5. Observe the waveform on the display. In normal condition, the square waveform
as shown in Figure 1-9 should be displayed.
Figure 1-9 Square Waveform Signal
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform "Probe
Compensation" introduced in the next section.
Compensation Signal Output Terminal
Ground Terminal

Table of Contents

Other manuals for Rigol MSO5074

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO5074 and is the answer not in the manual?

Rigol MSO5074 Specifications

General IconGeneral
Bandwidth70 MHz
Channels4
Timebase Accuracy± 2 ppm
Operating Temperature0 °C to +40 °C
Display800x480
Waveform Capture Rate500, 000 wfms/s
Vertical Resolution8 bits
InterfacesUSB Host, USB Device, LAN, HDMI
Input Impedance1 MΩ
Trigger TypesEdge, Slope, Pulse, Video, Pattern, Duration, Timeout, Window, Runt
Sample Rate8 GSa/s

Related product manuals