Chapter 1 Quick Start RIGOL
MSO5000 User Guide 1-9
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
make them match the input channels of the oscilloscope. Non-compensated or
poorly compensated probes may cause measurement inaccuracy or errors. The
probe compensation procedures are as follows:
1. Perform Step 1, 2, 3 and 4 specified in "Function Inspection".
2. Check the displayed waveforms and compare them with Figure 1-10.
Figure 1-10 Probe Compensation
3. Use a nonmetallic screwdriver to adjust the low-frequency compensation
adjustment hole on the probe until the waveform is displayed as "Perfectly
compensated" in the figure above.
WARNING
To avoid electric shock when using the probe, please make sure that the
insulated wire of the probe is in good condition. Do not touch the
metallic part of the probe when the probe is connected to high voltage
source.
Tip
The probe compensation signal can only be used for probe compensation
adjustment and cannot be used for calibration.
Over compensated Perfectly compensated Under compensated

Table of Contents

Other manuals for Rigol MSO5074

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO5074 and is the answer not in the manual?

Rigol MSO5074 Specifications

General IconGeneral
Bandwidth70 MHz
Channels4
Timebase Accuracy± 2 ppm
Operating Temperature0 °C to +40 °C
Display800x480
Waveform Capture Rate500, 000 wfms/s
Vertical Resolution8 bits
InterfacesUSB Host, USB Device, LAN, HDMI
Input Impedance1 MΩ
Trigger TypesEdge, Slope, Pulse, Video, Pattern, Duration, Timeout, Window, Runt
Sample Rate8 GSa/s

Related product manuals