CALCulate Subsystem FSE
1065.6016.12 6.56 E-16
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:MIDDle[:STATe] ON | OFF
This command switches on or off the measurement of the arithmetical mean between positive and
negative peak value.
Example: ":CALC:MARK:FUNC:SUMM:MIDD ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: VA
When the measurement is switched on, the summary marker is automatically activated (command
SUMMary:STATe set to ON). When it is switched off, the summary marker remains switched on
provided further measurements are selected. Otherwise the marker is switched off automatically.
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:MIDDle:RESult?
This command queries the result of the measurement of the arithmetical mean between positive and
negative peak value. Results of average calculation and peak hold are queried with commands
...:MIDDle:AVERage:RESult? and ...:MIDDle:PHOLd:RESult?.
Example: ":CALC:MARK:FUNC:SUMM:MIDD:RES? "
Features: *RST- value: -
SCPI: device-specific
Mode: VA
This command is only a query and thus has no *RST value assigned
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:MIDDle:AVERage:RESult?
This command queries the result of the measurement of the arithmetical mean between positive and
negative peak value if the average is calculated using the command
:CALCulate<1|2>:
MARKer<1 to 4>:FUNCtion:SUMMary:AVERage
.
Example: ":CALC:MARK:FUNC:SUMM:MIDD:AVER:RES? "
Features: *RST- value: -
SCPI: device-specific
Mode: VA
This command is only a query and thus has no *RST value assigned
:CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:MIDDle:PHOLd:RESult?
This command queries the result of the measurement of the arithmetical mean between positive and
negative peak value if the peak hold function is switched on using the command
:CALCulate<1|2>:MARKer<1 to4>:FUNCtion:SUMMary:PHOLd.
Example: ":CALC:MARK:FUNC:SUMM:MIDD:PHOL:RES? "
Features: *RST- value: -
SCPI: device-specific
Mode: VA