FSE CONFigure-Subsystem
1065.6016.12 6.75 E-16
:CONFigure:BURSt:POWer[:IMMediate]
This command selects measurement of the average carrier power of the base station or mobile.
Example: ":CONF:BURS:POW"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
This command is an event and thus has no query and no *RST value assigned.
:CONFigure:BURSt:POWer:COUNt 1 to 1000
This command sets the number of bursts used for the determination of measured values.
Example: ":CONF:BURS:POW:COUN 100"
Features: *RST value: 500 (GSM/DCS1800 phase I)
200 otherwise
SCPI: device-specific
Modes: BTS, MS
:CONFigure:BURSt:POWer:CONDition NORMal | EXTReme
This command defines the conditions for power measurement.
Example: ":CONF:BURS:POW:COND EXTR"
Features: *RST value: NORMal
SCPI: device-specific
Modes: BTS, MS
:CONFigure:BURSt:PTEMplate[:IMMediate]
This command selects measurement of power of the base station or mobile vs. time.
Example: ":CONF:BURS:PTEM"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
This command is an event and thus has no query and no *RST value assigned.
:CONFigure:BURSt:PTEMPlate:COUNt 1 to 1000
This command defines the number of bursts used for determining the measured value.
Example: ":CONF:BURS:PTEM:COUN 100"
Features: *RST value: 500 (GSM/DCS1800 phase I)
200 otherwise
SCPI: device-specific
Modes: BTS, MS

Table of Contents