EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 140

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FSE CONFigure-Subsystem
1065.6016.12 6.83 E-16
:CONFigure[:MS]:NETWork[:TYPE] PGSM | PGSM900 | EGSM |EGSM900 | DCS |GSM1800 |
PCS|GSM1900 | RGSM | RGSM900 | GSM850
This command selects the standard type according to which the mobile will work.
Example: ":CONF:NETW DCS"
Features: *RST value: GSM
SCPI: device-specific
Mode: MS
:CONFigure[:MS]:NETWork:PHASe 1 | 2 [,PLUS]
This command selects the phase of the standard according to which the mobile will work.
Example: ":CONF:NETW:PHAS 2"
Features: *RST value: 1
SCPI: device-specific
Mode: MS
:CONFigure[:MS]:TXSupp ON | OFF
This command defines that an additional carrier suppression of min. 20dB is taken into account for
the measurement. If there is already suppression, a more sensitive setting of the instrument is
selected.
Example: ":CONF:TXS ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: MS
For measurements in the RX-band the value is automatically set to ON.
:CONFigure[:MS]:PRESet
This command resets the parameters for the standard selected to their default values (DEFAULT
SETTINGS).
Example: ":CONF:PRES"
Features: *RST value: --
SCPI: device-specific
Mode: MS
This command is an event and has thus no query and no *RST value assigned.