FSE CONFigure-Subsystem
1065.6016.12 6.85 E-16
CONFigure:SPECtrum Subsystem
This subsystem provides the commands for configuring the measurements in the GSM BTS Analyzer
mode (FSE-K11) or in the GSM MS Analyzer mode (FSE-K10) used to determine the power of the
spectral contributions due to modulation and switching (modulation spectrum, transient spectrum).
COMMAND PARAMETERS UNIT COMMENT
CONFigure
:SPECtrum
:MODulation
[:IMMediate]
:COUNt
:RANGe
:TGATe
:SWITching
[:IMMediate]
:COUNt
--
<numeric_value>
ARFCn | TXBand | RXBand |
COMBined | DCSRx1800 |
G8Rxband | PCSRx1900
<Boolean>
--
<numeric_value>
--
--
--
--
no query, Option FSE-K11 or FSE-K10
Option FSE-K11 or FSE-K10
Option FSE-K11 or FSE-K10
Option FSE-K11
Option FSE-K11 or FSE-K10
no query
:CONFigure:SPECtrum:MODulation[:IMMediate]
This command selects measurement of the spectrum due to modulation.
Example: ":CONF:SPEC:MOD"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
This command is an event and has thus no query and no *RST value assigned.
:CONFigure:SPECtrum:MODulation:COUNt 1 to 1000
This command sets the number of bursts used for determining the average and maximum values.
Example: ":CONF:SPEC:MOD:COUN 100"
Features: *RST value: 500 (GSM/DCS1800 phase I)
200 otherwise
SCPI: device-specific
Modes: BTS, MS