FSE FETCh-Subsystem
1065.6016.12 6.115 E-16
FETCh:SPURious Subsystem
This subsystem provides the commands for reading out results of measurements in the GSM BTS
(FSE-K11) or GSM MS (FSE-K10) Analyzer mode which are used to determine spurious emissions,
without first restarting a new measurement.
COMMAND PARAMETERS UNIT COMMENT
FETCh Option FSE-K11 or FSE-K10
:SPURious
[:ALL]?
:STEP?
TXBand | OTXBand | RXBand |
IDLeband
--
query only
query only
:FETCh:SPURious[:ALL]? TXBand | OTXBand | RXBand | IDLeband
This command reads out the results of the measurement of spurious emissions of the base station
or mobile which is performed in the LIST mode.
Parameter: The result is output as a list of partial result strings separated by ’,’ in the following
(ASCII) format:
<Index>,<Freq1>,<Freq2>,<Level>,<Limit>, <Abs/Rel>,<Status> [,
<Index>,<Freq1>,<Freq2>,<Level>,<Limit>, <Abs/Rel>,<Status>]...
where the parts between ’[...]’ denote a partial result string that can be repeated n times.
<Index>: 0, if the partial result string characterizes a measurement range
current number <>0,
if the partial result string characterizes a single
limit excess.
<Freq1>: Start frequency of the measurement range or
frequency where the limit line is exceeded
<Freq2>: Start frequency of the measurement range or
frequency exceeding the measurement range. The value of
<Freq2> is equal to the value of <Freq1>, if either the
measurement is performed in the time domain or if the
partial result string contains a limit excess.
<Level>: Measured maximum level of the partial range or
measured level at the test point.
<Limit>: Limit in the partial range or at the test point
<Abs/Rel>: ABS <Level> and <Limit> are in absolute units (dBm)
REL <Level> and <Limit> are in absolute units (dBm)
<Status>: Result of the limit check in character data form:
PASSED no limit exceeded
FAILED limit exceeded
MARGIN margin exceeded