SENSe Subsystem FSE
1065.6016.12 6.174 E-16
:[SENSe<1|2>:]CORRection:TRANsducer:SCALing LINear | LOGarithmic
This command defines whether the frequency scaling of the transducer factor is linear or logarithmic.
Example: ":CORR:TRAN:SCAL LOG"
Features: *RST value: LINear
SCPI: device-specific
Mode: A
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.
:[SENSe<1|2>:]CORRection:TRANsducer:COMMent <string>
This command defines the comment for the selected transducer factor.
Example: ":CORR:TRAN:COMM ’FACTOR FOR ANTENNA’"
Features: *RST value ’’ (empty comment)
SCPI: device specific
Mode: A
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.
:[SENSe<1|2>:]CORRection:TRANsducer:DATA <freq>,<level>..
This command defines the test points for the selected transducer factor. The values are entered as a
series of frequency/level pairs. The frequencies must be in ascending order.
Example: ":CORR:TRAN:TRANsducer:DATA 1MHZ,-30,2MHZ,-40"
Features: *RST value: -
SCPI: device-specific
Mode: A
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent. The level values are
sent as dimensionless numbers; the unit is specified by means of the command
SENS:CORR:TRAN:UNIT.
:[SENSe<1|2>:]CORRection:TRANsducer[:STATe] ON | OFF
This command switches the selected transducer factor on or off.
Example: ":CORR:TRAN ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: A
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.