FSE SENSe Subsystem
1065.6016.12 6.185 E-16
:[SENSe<1|2>:]DDEMod:PRATe 1 | 2 | 4 | 8 | 16
This command determines the number of points per symbol.
Example: ":DDEM:PRAT 8"
Features: *RST value: 4
SCPI: device-specific
Mode: VA-D
:[SENSe<1|2>:]DDEMod:FILTer:MEASurement OFF | RCOSine | RRCosine | GAUSsian| B22 | B25 |
B44 | QFM | FM95 | QFR | FR95 | QRM | RM95 |
QRR | RR95 | A25Fm | EMES | EREF
This command selects the input filter for the test signal.
B22 Bessel 22
B25 Bessel 25
B44 Bessel 44
QFM or FM95 IS95-CDMA fm
QFR or FR95 IS95-CDMA fr
QRM or RM95 IS95-CDMA rm
QRR or RR95 IS95-CDMA rr
A25Fm APCO 25 fm
EMES EDGE mes
EREF EDGE ref
Example: ":DDEM:FILT:MEAS RCOS"
Features: *RST value: OFF
SCPI: device-specific
Mode: VA-D
:[SENSe<1|2>:]DDEMod:FILTer:REFerence RCOSine | RRCosine | GAUSsian| B22 | B25 | B44 |
QFM | FM95 | QFR | FR95 | QRM | RM95 | QRR | RR95
| A25Fm | EMES | EREF
This command selects the input filter for the reference signal.
Example: ":DDEM:FILT:REF RCOS"
Features: *RST value: GAUSsian
SCPI: device-specific
Mode: VA-D
:[SENSe<1|2>:]DDEMod:FILTer:ALPHa 0.2 to 1
This command determines the filter characteristic (ALPHA/BT). Step width is 0.05.
Example: ":DDEM:FILT:ALPH 0.5"
Features: *RST value: 0.3
SCPI: device-specific
Mode: VA-D