Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
108User Manual 1173.9286.02 ─ 15
Parameters:
<Shape> FC1
Limit line defined by the noise floor and 1 corner frequency.
FC2
Limit line defined by the noise floor and 2 corner frequencies.
FC3
Limit line defined by the noise floor and 3 corner frequencies.
FC4
Limit line defined by the noise floor and 4 corner frequencies.
FC5
Limit line defined by the noise floor and 5 corner frequencies.
NONE
No limit line.
*RST:
NONE
Example:
CALC:PNL:TYPE FC2
Selects a limit line with 2 corner frequencies.
Manual operation: See "Phase Noise Limit Line" on page 59
CALCulate:PNLimit:FAIL?
This command queries the limit check results for phase noise limit lines.
Return values:
<LimitCheck> 1
Limit check has passed.
0
Limit check has failed.
Example:
CALC:PNL:FAIL?
Queries the limit check result.
Usage: Query only
CALCulate:PNLimit:SLOPe<segment> <Slope>
This command defines the slope for a phase noise limit line segment.
Suffix:
<segment>
.
1...5
Selects the limit line segment.
Parameters:
<Slope> Level distance from the left border of the limit line segment to
the previous one.
*RST: 10
Default unit: dB
Example:
CALC:PNL:SLOP2 20
Defines a slope of 20 dB for the second limit line segment.
Using Limit Lines