Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
110User Manual 1173.9286.02 ─ 15
<k> Limit line
Parameters:
<Line> 1 to 8
number of the new limit line
<name>
String containing the name of the limit line.
Example:
CALC:LIM1:COPY 2
Copies limit line 1 to line 2.
CALC:LIM1:COPY 'FM2'
Copies limit line 1 to a new line named FM2.
Manual operation: See "New / Edit / Copy To" on page 62
CALCulate<n>:LIMit<k>:DELete
This command deletes a limit line.
Suffix:
<n>
.
Window
<k> Limit line
Usage: Event
Manual operation: See "Delete" on page 62
CALCulate<n>:LIMit<k>:FAIL?
This command queries the result of a limit check in the specified window.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 81.
Suffix:
<n>
.
Window
<k> Limit line
Return values:
<Result> 0
PASS
1
FAIL
Example:
INIT;*WAI
Starts a new sweep and waits for its end.
CALC2:LIM3:FAIL?
Queries the result of the check for limit line 3 in window 2.
Usage: Query only
Using Limit Lines