Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
111User Manual 1173.9286.02 ─ 15
Manual operation: See "Check Traces" on page 61
CALCulate<n>:LIMit<k>:NAME <Name>
This command selects a limit line that already exists or defines a name for a new limit
line.
Suffix:
<n>
.
Window
<k> Limit line
Parameters:
<Name> String containing the limit line name.
*RST: REM1 to REM8 for lines 1 to 8
Manual operation: See "New / Edit / Copy To" on page 62
See "Name" on page 63
CALCulate<n>:LIMit<k>:LOWer:STATe
<State>
This command turns a lower limit line on and off.
Before you can use the command, you have to select a limit line with CALCulate<n>:
LIMit<k>:NAME on page 111.
Suffix:
<n>
.
irrelevant
<k> Limit line
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See "Visible" on page 61
CALCulate<n>:LIMit<k>:STATe <State>
This command turns the limit check for a specific limit line on and off.
To query the limit check result, use CALCulate<n>:LIMit<k>:FAIL?.
Note that a new command exists to activate the limit check and define the trace to be
checked in one step (see CALCulate<n>:LIMit<k>:TRACe<t>:CHECk
on page 112).
Suffix:
<n>
.
irrelevant
<k> Limit line
Using Limit Lines