Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
112User Manual 1173.9286.02 ─ 15
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:LIM:STAT ON
Switches on the limit check for limit line 1.
Manual operation: See "Check Traces" on page 61
CALCulate<n>:LIMit<k>:TRACe <TraceNumber>
This command links a limit line to one or more traces.
Note that this command is maintained for compatibility reasons only. Limit lines no lon-
ger need to be assigned to a trace explicitly. The trace to be checked can be defined
directly (as a suffix) in the new command to activate the limit check (see
CALCulate<n>:LIMit<k>:TRACe<t>:CHECk on page 112).
Suffix:
<n>
.
Window
<k> Limit line
Example:
CALC:LIM2:TRAC 3
Assigns limit line 2 to trace 3.
Manual operation: See "Check Traces" on page 61
CALCulate<n>:LIMit<k>:TRACe<t>:CHECk <State>
This command turns the limit check for a specific trace on and off.
To query the limit check result, use CALCulate<n>:LIMit<k>:FAIL?.
Note that this command replaces the two commands from previous signal and spec-
trum analyzers (which are still supported, however):
CALCulate<n>:LIMit<k>:TRACe on page 112
CALCulate<n>:LIMit<k>:STATe on page 111
Suffix:
<n>
.
Window
<k> Limit line
<t> Trace
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:LIM3:TRAC2:CHEC ON
Switches on the limit check for limit line 3 on trace 2.
Using Limit Lines