Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
114User Manual 1173.9286.02 ─ 15
Parameters:
<LimitLinePoints> Variable number of x-axis values.
Note that the number of horizontal values has to be the same as
the number of vertical values set with CALCulate<n>:
LIMit<k>:LOWer[:DATA] or CALCulate<n>:LIMit<k>:
UPPer[:DATA]. If not, the R&S FSW either adds missing val-
ues or ignores surplus values.
The unit is Hz.
*RST: -
Manual operation: See "Data Points" on page 64
CALCulate<n>:LIMit<k>:CONTrol:SHIFt
<Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<k> Limit line
Parameters:
<Distance> Numeric value.
The unit depends on the scale of the x-axis.
Manual operation: See "Shift X" on page 64
CALCulate<n>:LIMit<k>:LOWer[:DATA] <LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Suffix:
<n>
.
irrelevant
<k> Limit line
Parameters:
<LimitLinePoints> Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with CALCulate<n>:
LIMit<k>:CONTrol[:DATA]. If not, the R&S FSW either adds
missing values or ignores surplus values.
*RST: Limit line state is OFF
Manual operation: See "Data Points" on page 64
CALCulate<n>:LIMit<k>:LOWer:MODE <Mode>
This command selects the vertical limit line scaling.
Using Limit Lines