3-3
Rosemount 3144P
HART
NOTE
If communication is initiated with an improved 3144P using a communicator
that only has a previous version of the transmitter device descriptors (DDs),
the communicator will display the following message:
NOTICE: Upgrade to the field communicator software to access new XMTR
functions.
Continue with old description?
YES: The communicator will communicate properly with the transmitter using
the existing transmitter DDs. However, new software features of the DD in the
communicator will not be accessible.
NO: The communicator will default to a generic transmitter functionality.
If YES is selected after the transmitter is configured to utilize the new features
of the improved transmitters (such as Dual Input configuration or one of the
added sensor input types–DIN Type L or DIN Type U), the user will
experience trouble communicating with the transmitter and will be prompted
to turn the communicator off. To prevent this from happening, either upgrade
the communicator to the latest DD or answer NO to the above question and
default to the generic transmitter functionality.