EasyManua.ls Logo

Rosemount 3152 User Manual

Rosemount 3152
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series Nuclear Pressure
Transmitters including the Rosemount
3152, 3153 and 3154
Question and Answer IconNeed help?

Do you have a question about the Rosemount 3152 and is the answer not in the manual?

Rosemount 3152 Specifications

General IconGeneral
Product TypePressure Transmitter
ManufacturerEmerson
Measurement TypePressure
DisplayOptional LCD
Output Signal4-20 mA
Temperature Range-40 to 85 °C
Process Connection1/2 NPT
Electrical Connection1/2" NPT conduit
Housing MaterialStainless steel

Summary

Rosemount Warranty and Limitations of Remedy

Product Qualification and Requirements

Details standards the transmitters are designed for and requirements for maintaining qualified status.

Rosemount 3150 Series Installation Guide

Installation Procedures

Steps for mechanical transmitter mounting, conduit installation, and electrical connections for the Rosemount 3150 Series.

Rosemount 3150 Series Calibration Procedures

Calibration Procedures

Step-by-step instructions for performing span and zero adjustments, including zero-based and elevated/suppressed zero calibrations.

Correcting for High Static Line Pressure Effects

Span Correction for DP Transmitters (Ranges 4-5)

Details the systematic span shift in Ranges 4 and 5 due to high static line pressure and how to correct it.

Method 1 for Span Correction

Provides Method 1 for adjusting transmitter output based on calculated corrected readings for high static line pressure.

Method 2 for Span Correction

Provides Method 2 for adjusting transmitter pressure input based on corrected readings for high static line pressure.

Rosemount 3150 Series Transmitter Operation

Transmitter Theory of Operation

Explains the operation of the 3150 series pressure transmitter using a block diagram and formulas.

Rosemount 3150 Series Maintenance and Troubleshooting

Electronics Assembly Checkout and Replacement

Procedure for checking and replacing the electronics assembly if defective.

Sensor Module Checkout and Replacement

Steps for checking the sensor module for defects and replacing it if necessary.

Disassembly and Reassembly Procedures

Detailed steps for disassembling and reassembling the Rosemount 3150 transmitter, including flange and housing.

Troubleshooting Common Issues

Guide to diagnose and correct common symptoms like high output, low output, or erratic output.

Rosemount 3150 Series Transmitter Spare Parts

Related product manuals