Initial Startup
4.7 Inserting Final Filter
t Remove sampling tubing and attach sterile final filter as described in the
operating instructions.
The final filter is supplied together with a bell assembly. To assemble the final filter,
proceed as follows:
t Press the final filter into the quick connector of the display | dispenser unit.
t Attach the bell assembly to the final filter and then confirm by pressing “OK”.
y After assembly of the sterile final filter, the display switches to the operating
mode.
y The timers for the cartridges, UV lamp (pro UV and VF versions only),
ultrafilters (pro UV and VF versions only) and sterile final filters are reset.
4.8 Rinsing the Final Filter
t Place a collection vessel (min. 6 liter capacity) with calibration marks under the
final filter. Then, open the water dispensing and rinse the final filter with 6 liters
of water. During rinsing, open the vent valve of the final filter so that enclosed air
can escape.
t Afterwards, attach the protective cap to the bell assembly.
Initial startup is complete.
Startup
Remove sampling tubing and
attach sterile final filter as
described in the operating
instructions
OK
Quick connector
Vent valve
Bell assembly