47
arium
®
pro Operating Manual
Technical Specifications
10. Technical Specifications
Dimensions W + H + D
Without head
350 + 492 + 451 mm
With head
435 + 501 + 476 mm
Empty weight approx. 17 kg
Operating weight approx. 27 kg
Clearance requirements Sides Left (door):
Right:
min. 150-230 mm
min. 80 mm
Front 100 mm minimum for opening the doors
Feed water requirements Water must be pretreated by:
Distillation
Deionization
Reverse osmosis
Distilled water
> 250 KO + cm (< 4 μS/cm)
RO water
TDS < 25 ppm CaCO
3
> 10 KO + cm (< 100 μS/cm)
Deionized water
TDS < 10 ppm CaCO
3
> 50 KO + cm (< 20 μS/cm)
All Turbidity < 1 NTU
Silica < 1000 ppb
TOC < 1000 ppb
Pressure From depressurized atmospheric tank inlet to a maximum inlet pressure of 6.9 bar (100 PSI)
Product water
18.2 MO
*cm at 25 °C
Bacteria < 1 CFU | mL
TOC at 50 ppb
feed water
< 5 ppb
1
@ 25 °C
Flow rate Up to 2.0 L/min at a minimum inlet feed water pressure of 2 bar (without final filter)
Up to 1.6 L/min at a minimum inlet feed water pressure of 2 bar (with final filter)
Ambient conditions Operation 5 °C – 30 °C; 80% relative humidity, non-condensing
Storage 5 °C – 45 °C; 80% relative humidity, non-condensing
Power supply 100 – 240 Vac (+/- 10%), 50 60 Hz, 80 W (max.), 130 VA (max.)
Connections Serial interface RS-232C
Character-coded
Plug connection 9-pin D-sub port
Transmission rate 19200baud
Data bits 8
Parity None (space for Sartorius data printer)
Stop bits 1
arium
®
Exchange interface For communication with arium
®
advance
Fuse
2 fuses 5+20 mm at power inlet, time lag, 250 V, 2 A/T
Installation Installation Category II (over-voltage) in accordance with IEC 664
Pollution degree 2 in accordance with IEC 664
Altitude limit: 2000 m
1
Values can vary depending on the quality of the feed water and the content of impurities,
as well as the type of cartridge used.