43
arium
®
pro Operating Manual
Care and Maintenance
t After you have replaced the sterile final filter, the timer must be reset manually.
To reset the timer, select “Sterile final filter timer” in the Care r menu.
t Confirm the process.
y The timer is now reset and the system will then switch to the operating mode.
The warning to change the sterile final filter is not deactivated until the timer is
reset.
7.4 Replacing Electrical Fuses
The arium
®
pro system has two fuses.
These are located in a fuse drawer on the top of the housing, on the right hand side,
next to the main power switch (see Chapter “1.2 Electrical Connections”, page 8).
Only trained and qualified personnel may replace defective fuses.
Always disconnect the arium
®
pro system from the AC power outlet before replacing
defective electrical fuses.
Remove the old fuses from their sockets and ensure that they are always replaced
with fuses of the same type and rating.
t Gently press the locking device of the fuse drawer towards the left and pull out
the fuse drawer.
t Replace the defective fuses.
t Push the fuse drawer back into the housing until it clicks into place.
Care
Depressurization
Sterile final filter timer
Care r Sterile final filter
Reset sterile final filter timer
Cancel