Care and Maintenance
t Place a vessel (1L) under the water outlet and start depressurization.
y The water pressure in the system is reduced. Water flows from the sampling
outlet.
t Switch off your arium
®
pro system at the main switch.
This message can only be cleared by switching off the device!
Perform the work pending.
Restarting after Depressurization
t Turn the arium
®
pro system back on at the power switch. The system starts up.
Next, connect the feed water tubing.
t Place a vessel (1L) under the water outlet and start the rinsing process.
y The system is rinsed for one minute. Your arium
®
then switches to the operating
mode.
Care r Depressurization
Collect water leaking from
dispensing unit in a vessel (1L)
and start depressurization
Start
Care r Depressurization
Depressurization
0.5 min
Cancel
Care r Depressurization
Switch off device
at the main switch
Care r Depressurization
Connect the feed water
tubing to the device
OK
Care r Rinsing
Place vessel (1L) under
the dispenser and start
dispensing water
Start
Care r Rinsing
Rinsing
1 min
Cancel