94 Cubis MSE Operating Instructions
Technical Specifications
Model-specific Data
Verified Models with EC Type Approval Certificate: Micro- and ultramicrobalances
Model MSE6.6S-0CE MSE2.7S-0CE MSE3.6P-0CE
Accuracy class* k k k
For verified models: EC Type Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 0�001 0�0001 0�001/0�002/0�005
Weighing capacity max* g 6�1 2�1 1�1/2�1/3�1
Calibration value e* mg 1 1 1
Min� load min* mg 0�1 0�01 0�1
Tare equalization range (subtractive) < 100 % from max� weighing capacity
Application range according to DIR* g 0�001 – 6�1 0�001 – 2�1 0�001 – 3�1
Optimal starting point
of the measuring range** mg 0�82 0�082 0�82
Typical stabilization time s < 5 < 7 < 5
Typical measurement time s < 8 < 10 < 8
External standard calibration value
(min� accuracy class) g 5 2 3
Application range (temperature) With “isoCAL” function: +5…+40°C |
Without “isoCAL” function: +15 … +25°C
Display result (depending on the set filter level) By selection of 1 of 4 optimized filter levels
Weighing pan size d mm 30 20 30
Weighing chamber height
(draft shield DM) mm 70 70 70
Protection Protected against dust and water
* DIR = Directive 2014/31/EU on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model