Technical Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Semi-microbalances 0�01 mg
Model MSE225S-0CE MSE225P-0CE MSE125P-0CE
Accuracy class* k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 0�01 0�01/0�02/0�05 0�01/0�1
Weighing capacity max* g 220 60/120/220 60/120
Calibration value e* mg 1 1 1
Min� load min* mg 1 1 1
Tare equalization range (subtractive) < 100% from max� weighing capacity
Application range according to DIR* g 0�001 – 220 0�001 – 220 0�001 – 120
Optimal starting point
of the measuring range** mg 8�2 8�2 8�2
Typical stabilization time s < 2 < 2 < 2
Typical measurement time s < 6 < 6 < 6
External standard calibration value g 200 (E2) 200 (E2) 100 (E2)
(min� accuracy class)
Application range (temperature) With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +15 to +25 °C
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display result (depending on the set filter level) 0�2 – 0�4
Weighing pan size (W + D) mm 85 + 85
Weighing chamber height (draft shield DU)
mm 261
Protection Protected against dust and water
Verified models with EC Type Approval Certificate: Analytical balances 0�1 mg
Model MSE524S- MSE524P- MSE324S- MSE224S- MSE324P- MSE124S-
0CE 0CE 0CE 0CE 0CE 0CE
Accuracy class* k k k k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 0�1 0�1/0�2/0�5 0�1 0�1 0�1/0�2/0�5 0,1
Weighing capacity max* g 520 120/240/520 320 220 80/160/320 120
Calibration value e* mg 1 1 1 1 1 1
Min� load min* mg 10 10 10 10 10 10
Tare equalization range (subtractive) < 100% from max� weighing capacity
Application range according to DIR* g 0�01–520 0�01–520 0�01–320 0�01–220 0�01–320 0�01–120
Optimal starting point
of the measuring range** mg 82 82 82 82 82 82
Typical stabilization time s < 1 < 1 < 1 < 1 < 1 < 1
Typical measurement time s < 3 < 3 < 3 < 3 < 3 < 3
External standard calibration value g 500 500 200+100 200 200 +100 100
(min� accuracy class) (E2 (E2) (E2) (E2) (E2) (E2
Application range (temperature) With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +15 to +25 °C
Display result (depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 85 + 85
Weighing chamber height (draft shield DU)
mm 261
Protection Protected against dust and water
* DIR = Directive 2014/31/EU on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model