Problem ... Causes ... Solution
The special – No keys have been – Press any key
code “b“ displayed pressed since the balance
does not go out was turned on
The weight – Unstable ambient – Set up the balance
readout changes conditions in another area
constantly – Too much vibration or – Access the menu to
the balance is exposed select the appropriate
to a draft code to adapte the balance
to the particular weighing
environment (see Part 2)
– A door of the draft – Make sure that all
shield chamber is not draft shield doors
completely closed are closed
(only applies to draft shields
with sliding doors)
– The round glass draft – Position the round
shield is not in place draft shield
– The draft shield base – To mount these components,
plate (4) or the cen- see “Getting Started”
tering disk (27) is on page 1–9
not mounted correctly
– The dust cover for – Press down on the adhesive
balances with a rectangular strips to attach the dust cover
pan is not attached correctly
and is touching
the weighing pan
– A foreign object – Remove foreign object
is caught between the
pan and the housing of
the balance
– The cover plate for the – Turn the cover plate to
port for the below-balance close the port for
weighing hanger below-balance weighing
has not been closed
– The sample does not
have a stable weight
(absorbs moisture or
evaporates)
1–29