Counting . . .
. .
.
A weight readout is stored as a reference sample quantity (factory setting: 10 pcs).
To change this value, see the description on page 4ā14. (The reference sample
quantity is displayed as long as the reference sample weight has not yet been
stored for counting.) While counting (piece count readout), you can press F to
update the reference sample for optimizing the accuracy (approx. double the
reference sample quantity). In the process, the reference weight is recalculated
based on the total reference sample weight and the piece count displayed.
Counting Small Parts Code 2 1 4
Menu codes used in the example:
Function Code
Counting 2 1 4
With data ID code 7 2 2
Ref. quantity + ref. weight 7 1 2
Application: counting of bulk quantities of items with the same weight
Step/(key ID)/key Readout Data output/printout
Place container on the pan +
22.65 g
c, t
0.00 g
Add 10 sample parts +
58.55 g
Store: (STO REF) F +
10 pcs
nRef + 10 pcs
wRef + 5.8550 g
To update ref. sample,
add more parts
(approx. 2x nRef) +
20 pcs
press (OPT) F +
5.8544 g
wRef + 5.8544 g
Fill container with
desired quantity +
286 pcs
p +
286 pcs
Qnt + 286 pcs
To obtain weight
readout: press w +
1676.66 g
p +
1676.66 g
N + 1676.66 g
The data ID codes mean:
nRef + 10 pcs reference sample quantity (piece count)
wRef + 5.8550 g reference weight
Qnt + 286 pcs calculated piece count
N +1676.66 g net weight
As an alternative, you can
select this output parameter:
Reference weight 7 1 3
4ā13