Calibration Functions on Standard Balances
Select the appropriate menu code to define the access status for each of
the following calibration functions, which are activated by holding down the tare
key for a few seconds:
External calibration C-E
Internal calibration C-I
Sensitivity test C-t
If the menu access switch is unlocked (“accessible” status indicated by -C- after
you have accessed the balance operating menu), the “external calibration”
function will be accessible even though you have set the menu code 1 9 2 for
“access denied.
External calibration Code
Accessible * 1 9 1
Access denied 1 9 2
Internal calibration Code
Accessible * 1 10 1
Access denied 1 10 2
Sensitivity test Code
Accessible * 1 11 1
Access denied 1 11 2
Quick CAL Using f
If your balance has a built-in calibration weight, you can select the “internal
calibration” function at any time by a touch of the f key (factory setting:
accessible). That’s why we call it “Quick CAL” for short.
You can also set a different menu code to change the function of the f key from
“internal calibration” to “sensitivity test.”
Both functions are accessible even if you select code 1 10 2 or 111 2, respectively.
Function of the f key Code
No function * 2 2 1
Internal calibration (“CAL I”) * 2 2 5
Sensitivity test (“CAL T”) 2 2 6
For codes 2 2 2 to 2 2 4, see Part 4, “Application Programs.
* = factory setting; depends on the balance model in some cases
213