Data Output at Defined Intervals
You can reduce the volume of data in the “auto print” mode by defining the interval
at which data will be output automatically. This auto print interval is based on the
number of display updates.
Auto print interval Code
1 display update * 6 3 1
2 display updates 6 3 2
5 display updates 6 3 3
10 display updates 6 3 4
20 display updates 6 3 5
50 display updates 6 3 6
100 display updates 6 3 7
Automatic Taring after Data Output
This convenient setting lets you checkweigh a series of samples or products
without having to unload the balance after each weighing procedure. This means
less work for you:
the sample remains on the pan after the weight readout has been printed or
transferred to an on-line computer
the balance is tared automatically after the weight readout has been printed or
transferred to an on-line computer
you simply load the next sample or part
Code
Data output without automatic taring * 6 4 1
Data output with automatic taring 6 4 2
* = factory setting