Sieve Analysis Code 2 1 5
Menu codes used in the example:
Function Code
Weighing in percent 2 1 5
With data ID code 7 2 2
Reference % and weight 7 1 2
Tare memory 2 2 2
Output individual tare weight *7 3 1
Application: sieve analysis of granular substances
Step/(key ID)/key Readout Data output/printout
Place set of sieves on pan +
505.18 g
c, t
0.00 g
Add sample to top
sieve on the stack +
160.58 g
Store wt. as 100%: +
100.0 %
pRef + 100 %
STO REF F Wxx% + 160.58 g
Remove stack of sieves; shake
xx.x %
Place empty container
on the pan and press t
0.0 %
Pour contents of sieve 1 +
12.6 %
into container: (TAR) f +
0.0 %
Prc1 + 12.6 %
Add contents of sieve 2: +
34.8 %
(
TAR) f +
0.0 %
Prc1 + 34.8 %
Add and store contents
of remaining sieves
The data ID codes mean:
pRef + 100 % reference percentage
Wxx% + 160.58 g net reference weight xx%
Prc1 + 12.6 % percentage of the sample
* = factory setting
As an alternative, you can select
this output parameter:
Reference weight 7 1 3
47

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Master series and is the answer not in the manual?

Sartorius Master series Specifications

General IconGeneral
BrandSartorius
ModelMaster series
CategoryScales
LanguageEnglish

Related product manuals