Checking Net Weights – Weight Readout – Code 2 2 3
To store a weight readout as the target weight, press f.
The target weight is indicated in the analog display as a bar graph. The last segment
of this bar graph ends exactly below the equal sign of the checkweighing indication.
If the weight of a product is “acceptable,” the last segment of the bar graph will
be within the tolerance range of the checkweighing indication. You can also read
the actual weight in digits.
Menu codes used in the example:
Function Code
Checking net weights 2 2 3
With data ID code 7 2 2
Tolerance + target weight 7 1 2
Application: weighing in the same amounts of sample
Step/(key ID)/key Readout Data output/printout
c; place sample
container on the pan; t
0.00 g
Weigh-in required
sample amount +
235.18 g
Store: (STO) f +
1.0 %
Lim + 1.0 %
Setp + 235.18 g
+
235.18 g
Remove container + ideal
sample; place another
container on the pan; t
0.00 g
Weigh-in sample using the
Checkweighing indication
analog display as a guide +
235.55 g
p +
235.55 g
N + 235.55 g
Continue as directed above
The data ID codes mean:
Lim + 1.0 % over/under limits in percent
Setp + 235.18 g target weight stored
N + 235.55 g net sample weight
As an alternative, you can select
this output parameter:
Target weight only 7 1 3
4–10