6–116–10
Specifications – Gold Series
1
)
Model GC 1201-G
Capacity/readability Grams 1,200 g/0.01 g
Troy ounces 38.5 ozt/0.0005 ozt
Pennyweights 771 dwt/0.005 dwt
Tare range (by subtraction) g –1,200
Reproducibility (Standard deviation) g ≤ ± 0.005
Linearity g ≤ ± 0.02
Response time (average) s ≤1.5
Shortest response time in the IQ-Mode s 0.8
Adaptation to ambient conditions and application requirements by selection of 1of 4 optimized filter levels
Display update (depends on the filter level selected) s 0.1– 0.4
Sensitivity drift within 283…303 K /K ≤ ±2 · 10
–6
Ambient temperature range K 273…313 (0°C…+40°C; 32°F…104°F)
Pan size mm/in 130/5.1 Ø
Net weight, approx. kg 3.6
AC power source/power requirements V~ via AC adapter, 230 or 115, +15%…–20%
Frequency Hz 48 – 60
Power consumption VA max.: 16; average: 8
Selectable weight units g, kg, ct, lb, oz, ozt, tlh, tls, tlt, gr, dwt and mg
Application programs mass unit conversion by toggling; weighing in percent; tare memory; net total, over/under checkweighing
(classification and filling); counting
Automatic zero tracking function (can be turned off by menu code) standard feature
Built-in interface RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
transmission rate: 150…19,200 baud, 1 or 2 stop bits, software/hardware handshake
Standard features/equipment supplied
Dust cover x
AC adapter x
Hanger for below-scale weighing x
1
) Cannot be upgraded to Master series balances but can be retrofitted with an IAC
(Integrated Applications Computer) for complex applications.