Toggling between Two Calculated Values after Backweighing and
“Standard Weighing”
After backweighing, you can press the w key to toggle between two calculated
values and the “standard weighing mode.” Once you have set the desired menu
codes, the following calculated values can be displayed:
Readout of the 1st Calculated Value
Calcuated value Code
For MC 210 S, For MC 5(-0CE),
MC 210 P, SC 2 (-0CE)
MC 410 S (-0CE) and MC 21 S
Residue in selected wt. unit 3 9 1* 3 8 1*
Residue in percent 3 9 2 3 8 2
Difference in selected wt. unit 3 9 3 3 8 3
Difference in percent 3 9 4 3 8 4
Readout of the 2nd Calculated Value
Calcuated value Code
For MC 210 S, For MC 5(-0CE),
MC 210 P, SC 2 (-0CE)
MC 410 S (-0CE) and MC 21 S
Residue in selected wt. unit 3 10 1 3 9 1
Residue in percent 3 10 2 3 9 2
Difference in selected wt. unit 3 10 3 3 9 3*
Difference in percent 3 10 4* 3 9 4
Calculation is done according to these formulas:
Residue (backweighed sample)
Residue in percent = ––––––––––––––––––––––––––––––––
·
100%
Net initial sample weight
Residue – initial sample weight
Difference in percent = ––––––––––––––––––––––––––––––
·
100%
Initial sample weight
* = factory setting