EasyManua.ls Logo

Sartorius MC 210 S - General Interface Specifications; Interface Type, Standards, and Transmission Rates

Sartorius MC 210 S
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Specifications
Type of interface Serial point-to-point connector
Operating mode Asynchronous, full-duplex
Standard V28, RS-232C, RS-423 specification
Handshake*) 2-wire interface: via software (XON/XOFF)
4-wire interface: via hardware handshake lines
Clear To Send (CTS) and Data Terminal Ready (DTR)
Transmission rates*) 150; 300; 600; 1,200; 2,400; 4,800;
9,600; 19,200 baud
Character coding 7-bit ASCII
Parity*) Mark**), space**), odd, even
Synchronization 1start bit; 1or 2 stop bits *)
Data output format*) 16 or 22 characters
of the balance
Character format*) – 1 start bit
– 7-bit ASCII
– 1 parity bit
– 1 or 2 stop bits
**) = can be changed by the user
**) = inaccessible for verified balances approved for use as legal
measuring instruments
4–2