EasyManua.ls Logo

Sartorius MC 210 S - Basic Balance Operation: On;Off, Self-Test, and Warmup; Turning Display On;Off and Self-Test; Automatic Warmup Procedure

Sartorius MC 210 S
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Turning the Display On and Off (Standby Mode)
Press the e key (31) to turn the display on or off.
Self-Test
After the balance has been turned on, an automatic
self-test of the balance’s electronic circuitry
is performed, and the draft shield closes automatically.
When a zero readout is displayed, the balance
is ready for operation.
Automatic Warmup of the Balance Electronics in
MC 210 S, MC 210 P and MC 410 S (-0CE), MC 21S
Once you have connected the balance to AC power
and turned on the power, the balance’s electronics
automatically begin to warmup. The timer for
warmup is set to 4 minutes. The remaining warmup
time in minutes and seconds is displayed in
a countdown mode.
When the remaining time displayed is less
than one minute, you can interrupt this warmup
procedure by pressing the e key to turn the
balance off and back on again. Following
the self-test and display of a zero readout,
the balance is ready to operate again.
After 4 minutes have elapsed, the balance displays
a zero readout and is then ready to operate.
* = including the Signatories of the Agreement on
the European Economic Area
Important Note Concerning Verified Balances
Approved for Use as Legal Measuring Instruments
in the E.U.*
For verified balances that have a verification
scale interval “e” which is greater than the scale
interval “d,” and a scale interval “d” 0.1 mg,
the last digit on the display is bordered.
1–20