Specifications
Model MC 210 S MC 210 P MC 410 S
Weighing range structure SuperRange PolyRange SuperRange
Weighing capacity g 210 60/110/210 410
Readability mg 0.01 0.01/0.02/0.05 0.1
Tare range (subtractive) g 210 210 410
Repeatability (standard deviation)* mg ±0.015/ ±0.02/0.4/ ±0.15
0.02/0.03 0.5
for weighing range g
0–50/50–100
0–50/50–100/
100210 100210
Linearity mg ±0.12 ±0.15 ±0.5
Linearity within 5 g mg ±0.02 ±0.02
Response time (average) s –––––––––––– 10 –––––––––––––––– 2.5 –––
Response time in the IQ-mode (0.01%) s –––––––––––– 2.5 –––––––––––––––– 1.5 –––
Adapt. to ambient condit. and applic. requirem.
––– By selection of 1of 4 optimized filter levels ––
Display update (dep. on the filter level select.) –––––––––––––––– 0.20.4 ––––––––––––––––
Allowable ambient operating temperature range
°C –––––––––––––––– +5...+40 ––––––––––––––––
Operating temperature range °C ––––––––––––––– +10 ... +30 –––––––––––––––
Sensitivity drift within +10...+30°C /°C ––––––––––––––––±1·10
6
––––––––––––––––
Pan size mm ––––––––––––––––––– 90 –––––––––––––––––––
Weighing chamber (Ø x H), approx. mm –––––––––––––––– 167 x 245 –––––––––––––––
Net weight, approx. kg –––––––––––––––––––12 –––––––––––––––––––
Power requirements V~ STNG6 power supply (AC adapter), 230 VAC or
115 VAC, –20%...+15%
Frequency Hz –––––––––––––––––4860 –––––––––––––––––
Power consumption, approx. VA ––––––––– maximum: 28; average:16 ––––––––
Selectable weight units g, kg, ct, lb, oz, ozt, tlh, tls, tlt, GN, dwt, mg, /lb,
tlc, mom, K, tol, bat, MS
Application programs Mass unit conversion; “EUREKA” air buoyancy
correction progr.; diff. weighing and
backweighing;
density determination; tare memory; net total;
weighing in percent; over/under checkweighing
(classification and filling); counting
Automatic zero tracking function,
can be turned off by menu code –––––––––––––––– Standard ––––––––––––––––
Built-in interface
RS-232 C-S/V24-V28, RS-423/V 10; 7-bit;
parity: even, mark, odd, space;
trans. rates: 150...19,200
baud;1or 2 stop bits;
software/hardware handshake
Standard features/equipment supplied
Dust cover/power supply
xx x
Hanger for below-balance weighing x x x
Self-calibrating function “isoCAL” x x x
ISO/GLP recording capability/numeric keys x x x
* = Standard deviation of the reproducibility according to DIN 1319, Part 3