ISO/GLP-compliant Printout or Record
Application:
Use of the balance in quality assurance systems and in areas subject to
compliance with ISO, GLP, GMP and EN, etc.
The balance can record all completed calibration/adjustment operations and
output data in compliance with the requirements of Good Laboratory Practice (GLP)
and ISO. The balance, interfaced with a data printer or a computer, creates
a document that records the date, time, serial number and model number, making
it possible to clearly trace data to the balance that generated it and the time
at which it was generated.
Select the ISO/GLP printout/record mode by setting the respective code
in the balance operating menu.
ISO/GLP-compliant printout/record mode Code
Off 8 10 1*
Only for calibration and linearization functions 8 10 2
Always on for calibration and weighing functions 8 10 3
The following menu code must be set in order to obtain
an ISO/GLP-compliant printout/record:
With data ID code 7 2 2
!
Important Note:
ISO/GLP-compliant printouts/records will not be output if the factory setting
7 2 1 “without data ID code” is selected. When the “auto print” data
output parameter (menu code 6 1 4 or 6 1 5) is set, only calibration and
linearization functions are printed/recorded.
* = factory setting
2–25

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MC 210 S and is the answer not in the manual?

Sartorius MC 210 S Specifications

General IconGeneral
BrandSartorius
ModelMC 210 S
CategoryScales
LanguageEnglish

Related product manuals