Automatic draft shield function Code
Off 8 8 1
Close draft shield – perform function – open draft shield 8 8 2
Close draft shield – perform function with stability symbol delay –
open draft shield 8 8 3
Close draft shield – perform function 8 8 4*
Close draft shield – perform function with stability symbol delay 8 8 5
Display Accuracy with the Draft Shield Automatically Opened or
Manually Adjusted
You can define the display accuracy for the automatically opened or manually
adjusted draft shield by changing the increments, also called “scale intervals”
(of the last numeral). The display increments possible are as follows: 1, 2, 5,10,
20, 50, etc.
To make this concept easier to understand, the six levels are designated as
“rounding factors” in the table summarizing the various menu code settings.
Display accuracy Code
Highest possible accuracy 8 9 1*
Rounding factor 2 8 9 2
Rounding factor 5 8 9 3
Rounding factor 10 8 9 4*
Rounding factor 20 8 9 5
Rounding factor 50 8 9 6
Rounding factor 100 8 9 7**
** = factory setting
** = does not apply to verified balances approved for use as legal
measuring instruments
2–24