Practical Example: Net Total
Menu code settings used in the example:
Function Code
Tare memory 2 2 2
Automatic data output 7 1 2
With data ID code 7 2 2
Printout of individual
components/tare weights 7 3 1*
Application: Weighing-in several components with additive storage, automatic
taring and output of the component weight
Step/Key ID/Key Readout Data Output
Place container on pan
122.650 mg
c, t
0.000 mg
Weigh in 1 st component +
1059.335 mg
Store: TAR - f
0.000 mg
Net
N1 + 1059.335 mg
Weigh in 2nd component +
1352.330 mg
Net
Store: TAR - f
0.000 mg
Net
N1 + 1352.330 mg
Weigh in and store additional
components +
xxxx.xxx mg
Net
N1 + xxxx.xxx mg
Finish weighing-in
procedure; 0; f
Output the total weight: p +
3181.985 mg
Net
N + 3181.985 mg
The data ID codes mean:
N1 + 1059.335 mg component weight
(net individual weight)
N + 3181.985 mg total amount weighed-in (net total weight)
* = factory setting
As an alternative, you can select this
code:
Printout of net total weight/
total tare weight 7 3 2
3–46