Practical Example: Checking Net Weights
Menu code settings used in the example:
Function Code
Checking net weights 2 2 3
Control by port lines for the
YRD10Z checkweighing display 8 7 1
Tolerance + target weight 7 1 2
With data ID code 7 2 2
Application: Weighing-in the same amounts of samples
Step/Key ID/Key Readout Data Output
c; place sample container
on pan; t
0.000 mg
Weigh in required
sample amount +
493.110 mg
Store: STO - f
1.0 %g
Lim + 1.0 %
Setp + 493.110 mg
+
493.110 mg
Remove container + ideal
sample; place another
container on pan; t
0.000 mg
Weigh in sample using Checkweighing
the analog bar graph indication
as a guide +
495.894 mg
p +
495.894 mg
N + 495.894 mg
Continue as directed above
The data ID codes mean:
Lim + 1.0 % over/under limits in percent
Setp + 493.110 mg target weight stored
N + 495.894 mg net sample weight
As an alternative, you can select
this code:
Target weight only 7 1 3