Linearization
Hold down t for 2 seconds
until the following appears:
Choose “L.I.” for
internal linearization
Tare
Internal:
Clear weighing pan
An acoustic signal indicates
the end of calibration
Display: 0.000 mg
External:
Clear weighing pan
Choose “C.E.” for
external calibration
Tare
Calibration
Internal when “CAL I”
is displayed:
Clear weighing pan
Tare
If necessary, enter
certified weight value using
the numeric keys:
Center required weight
on the weighing pan
An acoustic signal indicates
the end of linearization
Display: 0.000 mg
Hold down t for 2 sec.
until the following appears: