Stability Symbol Delay
This setting allows your balance to compensate for individual interfering factors
which slowly subside, such as turbulent air currents generated within the
weighing chamber.
Code
No delay 1 4 1
Short delay 1 4 2*
Long delay 1 4 3
Extremely long delay 1 4 4
Tare Parameter
You can define when the balance will perform the taring operation:
Code**
At any time 1 5 1
Not until the readout is stable 1 5 2*
Auto Zero Function
When this zero tracking function is activated, any changes off the zero readout
that are equal to a defined fraction of digits per second are automatically tared.
In other words, it ensures a stable zero.
Code
Auto Zero on 1 6 1*
Auto Zero off 1 6 2
* = factory setting
** = setting does not apply to verified balances approved for use as legal
measuring instruments