Using the Weight Set for Air Density Determination
– Steel Weight –
Use the numeric keys to enter the exact mass value
specified for the certified steel weight
(e.g. 4.999990) and store this value by
pressing the f key identified in the display by “STO.
If you press f, identified by “STO,” without
making a numeric entry, the display changes
to the density readout. Now you can
change it for a special steel weight. Enter
the mass value and press f to define
this density.
Center the steel weight on the weighing pan and close
the draft shield.
After the steel weight has been confirmed, the display
will show a zero readout. Now, remove the steel
weight from the weighing pan and close the
draft shield. Afterwards, if the value for the aluminum
weight “Al” is displayed, you can continue with
the procedure.
”Err 03: The effects of an external disturbance
can cause the error code “Err 03”
to be displayed briefly. If this occurs, restart
the air density determination procedure.