Practical Example: Residue Determination of Samples in Containers
Menu code settings used in the example:
Function Code
For MC 210 S,
For
MC 210 P,
MC 5 (-0CE),
MC 410 S (-0CE)
SC 2 (-0CE)
and MC 21 S
Differential
weighing and
backweighing 2 1 6 2 1 6
Consec. indiv.
weighing 3 7 2* 3 6 2*
Automatic
data output 7 1 2 7 1 2
With data
ID code 7 2 2 7 2 2
2 decimal
places for the
results in percent 3 6 3* 3 5 3
With tare
weighing 3 8 2 3 7 2*
1st calculated
value: residue
in wt. unit 3 9 1* 3 8 1*
2nd calculated
value: residue
in percent 310 2 3 9 2
Record/printout
with net initial
sample weight 7 5 3* 7 5 3*
Record/printout
with residue
in wt. unit
and percent 7 6 4 7 6 4
Application: Brief sample treatment, such as ashing (incineration).
A value is recorded/printed out automatically after backweighing.
As alternatives, you can select
these codes:
Different number
of decimals 3 6 x 3 5 x
Without tare
weighing 3 8 1 3 7 1
Different calculated
value 3 9 x 3 8 x
Different calculated
value 3 10 x 3 9 x
Other record
printout/output 7 5 x 7 5 x
Other record
printout/output 7 6 x 7 6 x
3–21