Practical Example: Residue Determination of Samples in Containers
Menu code settings used in the example:
Function Code
For MC 210 S,
For
MC 210 P,
MC 5 (-0CE),
MC 410 S (-0CE)
SC 2 (-0CE)
and MC 21 S
Differential 
weighing and
backweighing 2  1  6 2  1  6
Consec. indiv.
weighing 3  7  2* 3  6  2*
Automatic 
data output 7  1  2 7  1  2
With data 
ID code 7  2  2 7  2  2
2 decimal 
places for the 
results in percent 3  6  3* 3  5  3
With tare 
weighing 3  8  2 3  7  2*
1st calculated
value: residue 
in wt. unit 3  9  1* 3  8  1*
2nd calculated 
value: residue 
in percent 310  2 3  9  2
Record/printout 
with net initial 
sample weight 7  5  3* 7  5  3*
Record/printout 
with residue 
in wt. unit
and percent 7  6  4 7  6  4
Application: Brief sample treatment, such as ashing (incineration).  
A value is recorded/printed out automatically after backweighing.
As alternatives, you can select 
these codes:
Different number
of decimals 3   6  x 3  5  x
Without tare 
weighing 3   8  1 3  7  1
Different calculated
value 3   9  x 3  8  x
Different calculated 
value 3 10 x 3  9  x
Other record 
printout/output 7   5  x 7  5  x
Other record 
printout/output 7   6  x 7  6  x