72
SCAIME Manual : NU-eNod4D-ETH-E-0418_216710-E.docx
Reference
Reference value expressed in factory calibrated points for the
sensor(s) input control test. The value is automatically
determined and stored after executing the sensor input reference
command.
Result
Result of sensor(s) input control test expressed in 1/10 of user
weight unit. Its value is automatically determined and stored
after executing the sensor input control command. This test
result represents the weight difference between the reference
value and the current test value.
Result max.
tolerance
0.0 – 6553.5
The Sensor input control result variable is compared with the
Sensor input control result max. Tolerance parameter which is
expressed in 1/10 of user weight unit and has a default value of
30. If the sensor input control result value is greater than or equal
to “Sensor input control result max. Tolerance” then an error is
reported
Sensor input control command will cause eNod4 to handle special
test on sensor input and to deliver a test result. This command
must not be realized when any process cycle that use weight is in
progress.
Sensor input reference command will cause eNod4 to handle
special sequence to acquire sensor input control reference value
of the load cell sensor input. This command must be performed
before using the Sensor input control command and each time
the weight measurement chain (sensor, connection, calibration)
is modified. This command must not be realized when any
process cycle that use weight is in progress.