SEAWARD PRIMETEST SERIES USER MANUAL
TESTED. TRUSTED… WORLD WIDE Page 13 of 77
5.4 PrimeTest Battery Status Indication
While the PrimeTest is powered on there are periodic checks of the batteries. The PrimeTest will show the
status of the batteries.
Note: The battery symbols are not displayed in real time and may take a few minutes to update
This symbol indicates that the batteries are at 100% capacity.
There are several symbols which will display the current battery voltage. When these icons are
displayed the batteries are still good.
This symbol indicates that the batteries are low. Although tests performed with the batteries in this
state are still valid, all test types may not complete their intended duration. The batteries should
be recharged.
This symbol indicates that the batteries are discharged. The PrimeTest will switch itself off after a
short period of time. The batteries must be recharged.
This symbol indicates that the PrimeTest is bulk charging batteries at the full charge current.
This symbol indicates that the PrimeTest is trickle charging the batteries with reduced charge
current. This is known as the Top Up Charge.
This symbol indicates that the PrimeTest has detected a fault with the battery or battery charger
circuit. Unplug the PrimeTest from the mains supply and wait 2 minutes before reconnecting the
mains. If the fault persists then return the PrimeTest for service.
5.5 Charging the Battery Pack
The battery pack will be charged whenever the PrimeTest is connected to the Mains Supply regardless of
whether the instrument is switched ON or OFF.
The typical charging current is set to 500mA but this may vary as the instrument also includes pre-charge
and top up charge modes.
When no tests are being performed the battery pack should be fully recharged after 7 hours.
5.6 Replacing the Battery Pack
Before opening the PrimeTest Series battery compartment ensure that all test leads and
accessories are disconnected.
Power off the instrument.
Disconnect all test leads and accessories.
Position the instrument face down and remove the screw holding the battery compartment cover in
place.
Remove the battery compartment cover.
Remove the battery pack from the compartment and unplug the multiway connector.
Connect the multiway connector of the new battery to the connector in the battery compartment; place
the new battery pack into the compartment.
Relocate the battery compartment cover and fasten in position with the battery compartment screw.
Only use a replacement battery pack that has been supplied by Seaward or a Seaward approved
distributor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Seaward PrimeTest Elite+ and is the answer not in the manual?

Seaward PrimeTest Elite+ Specifications

General IconGeneral
BrandSeaward
ModelPrimeTest Elite+
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals