16 Useful Information
16.1 Factory Set Test Sequences
Test
Hot
Key
Elite+
Hot
Key
Pro
Visual
Earth
Continuity
Insulation
PE Conductor
Current
Touch Current
Substitute
(Alternative)
Leakage
Wiring
RCD
Trip
Time
Name
Duration
(s)
Limit
(Ω)
Duration
(s)
Voltage
Limit
(MΩ)
Duration
(s)
Limit
(mA)
Duration
(s)
Limit
(mA)
Duration
(s)
Limit
(mA)
Check
Limit
(ms)
Class 1 Earthed
500Vdc
A
A
Yes
2
1.0
5
500
1
Class 1 Leakage
230Vac
B
B
Yes
2
1.0
5
5.0
Class 2 Double ins
500Vdc
C
C
Yes
5
500
1
Class 2 Leakage
230Vac
D
D
Yes
5
1.0
Lead
E
E
Yes
2
1.0
5
500
1
Yes
Lead-RCD
F
F
Yes
2
1.0
5
1.0
Yes
300
Powerboard
G
G
Yes
2
1.0
5
500
1
Yes
Powerboard-RCD
H
H
Yes
2
1.0
5
1.0
300
Powerboard-Surge
250Vdc
I
I
Yes
2
1.0
5
250
1
Yes
M/Wave Emission
J
-
Yes
Visual Log
K
J
Yes
Asset Log
L
K
Yes
Leakage using PAC-
TPL
M
L
Yes
2
1.0
2
5
Yes
3 Phase insulation
using TPI
N
M
Yes
2
1.0
5
500
1

Table of Contents