SEAWARD PRIMETEST SERIES USER MANUAL
TESTED. TRUSTED… WORLD WIDE Page 6 of 77
1 User Notes
This instrument and its operating instructions are intended for use by adequately trained personnel.
Appropriate PPE should be worn at all times when using any PrimeTest unit. Electrically insulated
gloves should be used when performing tasks on live circuits above 50V. Arc-flash PPE selection
should be based on an incident energy analysis.
The following symbols are used in these operating instructions.
Warning of electrical danger!
Indicates instructions must be followed to avoid danger to persons.
Important, follow the documentation!
This symbol indicates that the operating instructions must be adhered to in order to avoid
danger.
Due to a policy of continuous development SEAWARD Electronic Limited reserves the right to alter the
equipment specification and description outlined in this publication without prior notice and no part of this
publication shall be deemed to be part of any contract for the equipment unless specifically referred to as an
inclusion within such contract.
2 Safety Notes
This PrimeTest unit has been built and tested in accordance with:
BS EN 61010-1, 2-030.
BS EN 61557 part 1, 2, 4 and 10.
To ensure safe operation of the unit, all notes and warnings in these instructions must be observed at all
times.
If the PrimeTest is used in a manner not specified by these operating instruction
then the protection provided may be impaired
Always ensure that the circuit or appliance under test is electrically isolated.
Do not connect the PrimeTest to electric circuits with nominal voltage greater than
CAT II 300 V AC/DC.
The PrimeTest and all associated cables and leads must be checked for signs of
damage before equipment is operated. Do not use if there are signs of damage.
Only use the correct leads supplied with the PrimeTest instrument.
Do not touch test probes beyond the hand barrier on the test probe.
The PrimeTest may apply high voltage or mains power to the appliance under test.
Do not touch conductive parts of the appliance while tests are active.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Seaward PrimeTest Elite+ and is the answer not in the manual?

Seaward PrimeTest Elite+ Specifications

General IconGeneral
BrandSeaward
ModelPrimeTest Elite+
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals