BASIC
- 31 -
SECOMAM - Polyphotometer UV-Vis BASIC V1.11
Date: 06/06/1997 Hour: 09H27
User:
Analysis: ALBUMIN 37°C
S a m p l e A b s . C o n c e n t r a t i o n
S t a t u s
R e a g e n t
B l .
0 . 0 0 0
0 . 0 0 g / l
C o n t r o l 0 . 9 9 0
9 . 9 0 g / l P B
C o n t r o l 0 . 9 9 0
9 . 9 0 g / l P B
C o n t r o l P b L e 0 7 / 0 6 / 9 7
N o r m a l v a l u e : 3 8 . 0 0 - 5 4 . 0 0 g / l
Status:
>
: Sample concentration superior to upper limit.
< : Sample concentration inferior to lower limit..
PB : Problem on control.
OK : satisfactory control.
Q : Appears next to analysis result when control is programmed but not run.
! : Appears next to analysis result when the control is out of tolerance range.
L : The result in concentration is no longer within the linearity limit of the analysis.
− : The result in concentration is negative.
Upon modifying analysis method, memorized results are erased.
3
day